Users Manual
Specifications
33
IEC Lead Test
Earth bond, limit ....................................................... 0.10
Test current .............................................................10 A ac
Insulation resistance, limit ....................................... 2.0 M
Test voltage......................................................... 500 V DC
Fuse test and polarity check
PELV Test
Display ............................................. “> PEL” indicator only
Accuracy at 50 Hz ..................................± ( 2% + 3 counts)
Overload protection ............................................. 300 Vrms
Warning threshold ................................................. 25 Vrms
Test Limits for PASS result
R
PE
200 mA........................................................... < 0.10
R
PE
25 A................................................................ < 0.10
R
ISO
............................................................. > 1 M Class I
> 2 M Class II
I
SUB
...................................................................... < 3.5 mA
I
PE
...................................................................... < 0.75 mA
I
TOUCH
................................................................ > 0.25 mA
IEC Lead Test:
R
PE
........................................................................ < 0.10
R
ISO
......................................................................... > 2 M
Variation Factor Errors
Variation
Factor
Designation % Variation
Error
Position E1 0.0%
Supply Voltage E2 5.0%
Temperature E3 5.5%
Current
Consumption
E4 1.5%
Magnetic Fields E5 2.5%
Impedance E6 1.0%
Capacitance E7 2.0%
Current
Waveshape
E8 1.0%