Users Manual
32
Substitute Leakage Current Test (I
SUB
)
Operational Error Measurement Range ... 0.25 to 19.00 mA
Operational Error ......................................................... 10%
Accuracy ................................................. ± (5% + 5 counts)
Display Range ......................................... 0 to 19.99 mA ac
Resolution..............................................................0.01 mA
Test Voltage................................................. 35 V ac ± 20%
(at nominal mains input voltage)
Touch Current Test (I
TOUCH
)
Operational Error Measurement Range....... 0.1 to 1.99 mA
Operational Error ........................................................ 6.0%
Accuracy ................................................. ± (4% + 2 counts)
Display Range ........................................... 0 to 1.99 mA ac
Resolution..............................................................0.01 mA
Internal Resistance (via probe) ....................................2 k
Measuring method .....................................................Probe
The appliance under test is energized at mains potential.
Load/ Leakage Test: Load Current
Display Range ......................................................0 to 13 A
Accuracy ................................................. ± (4% + 2 counts)
Resolution...................................................................0.1 A
The appliance under test is energized at mains potential.
Load/Leakage Test: Load Power
Display Range ................................................. 0 to 999 VA
1.0 kVA to 3.2 kVA
Accuracy ................................................. ± (5% + 3 counts)
Resolution ............................................ 1 VA (0 to 999 VA)
0.1 kVA (1.0 kVA to 3.2 kVA)
The appliance under test is energized at mains potential.
Load/Leakage Test: Leakage Current (I
PE
)
Operational Error Measurement Range ... 0.25 to 19.00 mA
Operational error....................................................... 12.0%
Accuracy ................................................. ± (4% + 5 counts)
Display Range .........................................0.25 to 19.99 mA
Resolution..............................................................0.01 mA
The appliance under test is energized at mains potential.