Users Manual
Operating Features
7
No Annunciator Meaning
1
Caution! Risk of Danger. Refer to Manual.
2
Please check the mains polarity.
3
MULTI RESULTS Multiple test results will be displayed in succession.
4
PRINT Printing results.
5
Fluke SP1000 mini printer is connected.
6
INSULATION Insulation test.
R
ISO
<2.0 MΩ CII
<1.0 MΩ CI
Lights up if the recommended insulation resistance (R
ISO
) limit is exceeded
(Class II <2.0 M , Class I <1.0 M).
I
SUBSTITUTE
Substitute leakage test.
EARTH BOND
200 mA
25 A
Earth Bond test 200 mA or 25 A.
R
PE
>0.10Ω
+LEAD?
Lights up if the recommended protective earth conductor resistance (R
PE
) limit is
exceeded. Notice that a long connection lead can add extra resistance.
I
TOUCH
Touch current test
IEC LEAD
IEC lead test.
LOAD PE LEAK
Load/Leakage Current test
Limit
The test result should be checked against the limit allowed for the appliance.