Fluke 6200 - Bond Test Procedure and Results

Fluke 6200
39 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Users Manual
Testing Appliances
15
2
BOND
200mA
BOND
25A
Start the 200 mA test or the 25 A test:
Single test - press momentarily
Continuous test - hold down > 2 seconds
The display shows the test progress, see
the table below.
3
During the measurement flex the flexible
cord along its length to help find any broken
conductors or poor quality joints.
4
BOND
200mA
BOND
25A
Continuous test only: stop the test.
5
When the test is finished remove the earth
bond lead from the appliance.
6
MEM
Store the test result, if required.
Note
If a double beep sounds the earth bond test lead
has not been zeroed (no
symbol on the LCD).
You must zero the test lead, see page 10.
Figure 4. Bond Test Connections
DISPLAY EXPLANATION
EARTH BOND R
PE
200 mA 25 A
The 200 mA or the 25 A earth bond test
has been selected.
seconds The test time is counted down.
The bond test has been zeroed.
Ω R
PE
is 0.03 .
> Ω R
PE
overrange.
Limit > 0.10
+ LEAD?
R
PE
may have exceeded the
recommended limit, possibly because of
the length of the supply lead.