Users Manual
Testing Appliances
15
2
BOND
200mA
BOND
25A
Start the 200 mA test or the 25 A test:
Single test - press momentarily
Continuous test - hold down > 2 seconds
The display shows the test progress, see
the table below.
3
During the measurement flex the flexible
cord along its length to help find any broken
conductors or poor quality joints.
4
BOND
200mA
BOND
25A
Continuous test only: stop the test.
5
When the test is finished remove the earth
bond lead from the appliance.
6
MEM
Store the test result, if required.
Note
If a double beep sounds the earth bond test lead
has not been zeroed (no
∅
symbol on the LCD).
You must zero the test lead, see page 10.
Figure 4. Bond Test Connections
DISPLAY EXPLANATION
EARTH BOND R
PE
200 mA 25 A
The 200 mA or the 25 A earth bond test
has been selected.
seconds The test time is counted down.
∅
The bond test has been zeroed.
Ω R
PE
is 0.03 .
> Ω R
PE
overrange.
Limit > 0.10 Ω
+ LEAD?
R
PE
may have exceeded the
recommended limit, possibly because of
the length of the supply lead.