Fluke 6200 - Substitute Leakage Current Test Procedure

Fluke 6200
39 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6200
Users Manual
18
2
I
SUB
Start the test:
Single test - press momentarily
Continuous test - hold down > 2 seconds
The display shows the test progress, see the
table below.
3
I
SUB
Continuous test only: stop the test.
4
MEM
Store the test result, if required.
5
For Class II continue the test for all exposed
metal parts on the appliance.
Figure 6. Substitute Leakage Current Connections
DISPLAY EXPLANATION
I
SUBSTITUTE
The substitute leakage current test has
been selected.
seconds The test time is counted down.
mA
> mA
I
SUBSTITUTE
is 0.13 mA.
I
SUBSTITUTE
overrange.
Limit
The acceptable test limit may have
been exceeded. Refer to standards
and/or guidance materials.