6200
Users Manual
18
2
I
SUB
Start the test:
Single test - press momentarily
Continuous test - hold down > 2 seconds
The display shows the test progress, see the
table below.
3
I
SUB
Continuous test only: stop the test.
4
MEM
Store the test result, if required.
5
For Class II continue the test for all exposed
metal parts on the appliance.
Figure 6. Substitute Leakage Current Connections
DISPLAY EXPLANATION
I
SUBSTITUTE
The substitute leakage current test has
been selected.
seconds The test time is counted down.
mA
> mA
I
SUBSTITUTE
is 0.13 mA.
I
SUBSTITUTE
overrange.
Limit
The acceptable test limit may have
been exceeded. Refer to standards
and/or guidance materials.