EasyManua.ls Logo

Fluke 702 - DC Current Measurement; Resistance Measurement

Fluke 702
120 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Documenting Process Calibrator
Specifications
89
DC Current Measurement
RANGE RESOLUTION
MODEL 702
% OF READING +% OF FULL SCALE
MODEL 701
% OF READING +% OF FULL SCALE
1 YEAR 2 YEAR 1 YEAR 2 YEAR
30 mA 1 µA 0.025% + 0.025% 0.025% + 0.03% 0.025% +0.025% 0.025% + 0.03%
110 mA 10 µA 0.05% + 0.05% 0.07% + 0.07% 0.05% + 0.05% 0.1% + 0.07%
Temperature Coefficient: (0.001% of rdg. + 0.002% f.s.)/°C in the ranges -10 to 18°C and 28 to 50°C
Common Mode Error: 0.01% f.s./(Common Mode Volt)
Maximum Input Voltage: 30V dc
Resistance Measurement
RANGE RESOLUTION
MODEL 702
% OF READING + Ohms
MODEL 701
% OF READING + Ohms
1 YEAR 2 YEAR 1 YEAR 2 YEAR
11 0.001 0.05% + 0.05 0.075% + 0.05 0.1% + 0.05 0.15% + 0.05
110 0.01 0.05% + 0.05 0.075% + 0.05 0.1% + 0.05 0.15% + 0.05
1.1 k 0.1 0.05% + 0.5 0.075% + 0.5 0.1% + 0.5 0.15% + 0.5
11 k 1 0.1% + 10 0.1% + 10 0.1% + 10 0.15% + 10
Temperature Coefficient: (0.01% f.s. + 2 mΩ) /°C in the ranges -10 to 18°C and 28 to 50°C
Common Mode Error: 0.005% f.s/(Common Mode Volt)
Maximum Input Voltage: 30V dc

Table of Contents

Other manuals for Fluke 702

Related product manuals