EasyManua.ls Logo

Fluke 702 - Loop Power Supply

Fluke 702
120 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Documenting Process Calibrator
Specifications
99
Temperature, Resistance Temperature Detectors (cont)
100 Pt(3916)
-200 to -190 0.3 0.4 0.3 0.4
-190 to 0 0.3 0.4 0.1 0.2
0 to 630 0.5 0.8 0.2 0.4
200, 500, 1000 Pt(385)
-200 to 0 0.3 0.5 0.1 0.2
0 to 400 0.5 0.8 0.2 0.4
400 to 630 0.8 1.0 0.4 0.5
Sensor inaccuracies not included
Resolution: 0.1°C
Temperature Coefficient: 0.02°C/°C in the ranges -10 to 18°C and 28 to 50°C
Maximum Input Voltage: 30V
For two and three-wire RTD measurements, add 0.4°C to the specifications.
Loop Power Supply
MODELS 702 AND 701
SETTING 1 YEAR 2 YEAR
24 Volt 5% 5%
28 Volt 5% 5%
Short circuit protected
Maximum Current: 22 mA
Maximum Input Voltage: 30V dc

Table of Contents

Other manuals for Fluke 702

Related product manuals