Documenting Process Calibrator
Specifications
93
Resistance Sourcing
RANGES RESOLUTION
MODEL 702
% OUTPUT + OHMS
MODEL 701
% OUTPUT + OHMS
1 YEAR 2 YEAR 1 YEAR 2 YEAR
11.000 1 m 0.01% + 0.02 0.02% + 0.02 0.02% + 0.02
0.03% + 0.02
110.00 10 m 0.01% + 0.04 0.02% + 0.04 0.02% + 0.04
0.03% + 0.04
1.1000 k 100 m 0.02% + 0. 5 0.03% + 0.5 0.05% + 0.5
0.1% + 0.5
11.000 k 1 0.03% + 5 0.04% + 5 0.1% + 5
0.15% + 5
Temperature Coefficient: (0.01% of f.s.)/°C in the ranges -10 to 18°C and 28 to 50°C
Maximum/Minimum Current through Source Resistance:
11 Range: 3 mA dc max, 0.1 mA dc min
110 Range: 3 mA dc max, 0.1 mA dc min
1.1 k Range: 3 mA dc max, 0.01 mA dc min
11 k Range: 1 mA dc max, 0.01 mA dc min
Common Mode Error: 0.008% f.s/(Common Mode Volt)
Maximum Input Voltage: 30V dc