Documenting Process Calibrator
Simultaneous Measure/Source
61
Table 8. Simultaneous MEASURE/SOURCE Functions with Loop Power Enabled
MEASURE FUNCTION
SOURCE FUNCTION
dc V mA Freq TC 2W RTD Pressure
dc V
mA
ac V
Frequency (20 Hz)
TC
Pressure ••
The two softkeys displayed in MEASURE/SOURCE
mode are for use in calibrating a process instrument.
Instructions for calibrating a process instrument are
provided next.
As Found, which starts up a calibration routine
to obtain and record as found data.
Adjust, for making adjustments to a transmitter if
required.