Precision Loop Calibrator
Calibrator Setup Menu
19
Valve Test
The Valve Test function turns on or off the valve test
feature. See Figure 13.
gzx25.eps
Figure 13. Valve Test Enable Screen
The valve test verifies correct operation of valves. In
valve test, the output can be stepped to these values:
3.8 mA
4.0 mA
4.2 mA
8.0 mA
12.0 mA
16.0 mA
19.8 mA
20.0 mA
20.2 mA
The valve test mA values are not affected by the mA span
setting.
1. Use the Main menu to set the Product to source or
simulate current.
2. If valve test is not enabled, use the menu to enable it.
3. Push or to step the output to verify the
proper valve operation.
4. Use the menu to disable valve test when done.