EasyManua.ls Logo

Fluke 709 - Maintenance

Fluke 709
62 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
709/709H
Users Manual
44
The data items shown are:
The top line indicates is logging is in progress
(Logging) or is stopped (Stopped). Logging stops
automatically when storage is full or before the
Product is turned off when the battery reaches its low
voltage automatic shutoff limit of 5.6 V.
Interval is the item previously selected.
Elapsed is the time since the log was started,
updated each time a new sample is saved.
Records used is the total number used to date for all
sessions, updated each time a new sample is saved.
Records free is the total number that remain
unused, updated each time a new sample is saved.
709H is the present measurement, updated as often
as possible.
PV mA is the last HART device measurement,
updated as often as possible.
Maintenance
Warning
For safe operation and maintenance of the
Product:
Repair the Product before use if the
battery leaks.
Be sure that the battery polarity is
correct to prevent battery leakage.
Remove the input signals before you
clean the Product.
Use only specified replacement parts.
Have an approved technician repair the
Product.

Other manuals for Fluke 709

Related product manuals