Write LRV and URV Values
33
Write LRV and URV Values
Note
For the screens in this section, tag names can
be truncated to fit on one line if necessary.
flashes in the upper-right corner to show a live
connection.
If the HART write commands are not active, these
functions are not available and an error message is
shown instead of the screen in Figure 32.
gzx41.eps
Figure 32. Write LRV and URV Screen
Use the selection knob to choose the necessary function.
Push the selection knob to choose the highlighted action
and go to the corresponding screen. These screens are
explained in the “Write LRV” and “Write URV” sections.
Push to go to the Function Select menu.
Write LRV
The Product warns you to change the loop to MANUAL
before you proceed. Push the selection knob to proceed.
Push to go to the Write LRV and Write URV menu.
The present LRV value and units are shown. See
Figure 33.
gzx42.eps
Figure 33. Write LRV Screen
1. Push the selection knob to move the decade cursor.
2. Turn the selection knob to increment or decrement
the value in steps indicated by the selected decade.
3. Push and hold the knob to send the new value to the
HART device. An error is shown if the HART device
rejects the value.