INTRODUCTION
SPECIFICATIONS
1-4
Table 1-2.
BOSOA
Specifications
(cont)
DC Current
RANGE
RESOLUTION
ACCURACY
for 1-Year
BURDEN VOLTAGE
200µA
0.01
µA
2mA
0.1
µA
±(0.3% of reading+ 2 digits 0.3V
max
20mA
1
µA
200
mA
10µA
2000
mA
100µA
0.9V
max
OVERLOAD PROTECTION
(ac/dc) .
. . . . . 2A/250V fuse in series with
3A/600V
fuse (for high energy sources)
AC Current (True
RMS Responding, AC Coupled)
INPUT
RESOLUTION
RANGE
20
Hz**
45
Hz
2 kHz
10
kHz
20
kHz
BURDEN
CURRENT
VOLTAGE
10µA
·
200µA
0.01
µA
200µA
I
100µA-2mA
0.1
µA
2mA
I
I
1
mA-20mA
1
µA
20mA
2%+10
1%+10
2%+10
0.3V
rms
max
I
10mA
- 200
mA
10µA
200mA
100
mA
· 2000
mA
100µA
2000mA
0.9V
rms
max
**Typically
3
to
5 digits
of
rattle
will
be
observed
at
full
scale
at
20
Hz.
CREST FACTOR
RANGE .
. . . . . . . . . . . . . . Waveforms with a Peak/RMS ratio
of
1:1
to
3:1
at full
scale
Resistance
RANGE
RESOLUTION ACCURACY
for
1-Year
FULL
SCALE VOLTAGE
ACROSS
UNKNOWN RESISTANCE
20on
o.orn
±(0.1% reading+ 2 digits+
.02Q)
.19V
2kQ
o.1n
1.2V
20kD.
m
.2V
200 kQ
1on
±(.05%
of
reading+ 2 digits)
2V
2000
kQ
1oon
.2V
20MQ
1kQ
±(0.25% reading+ 3 digits)
2V
OVERLOAD PROTECTION
............
. 500V
dc/ac rms on all
ranges
OPEN CIRCUIT VOLTAGE
..........•...
RESPONSE TIME
.....................
.
(TO
RATED
ACCURACY)
DIODE TEST
....
Less than 3.5V on all
ranges
10
seconds maximum on
20
MO
range; 2 seconds maximum on
all
other ranges
These three ranges have enough
voltage
to
turn on silicon
junctions
to check for proper forward-to-back resistance. The 2
kQ
range is
preferred and is marked with a larger diode symbol on the front
panel
of
the instrument.
Thethree
non-diode test ranges will
not
turn
on silicon junctions so in-circuit resistance measurements can be
made with these three ranges