ACCESSORIES
PROBE ACCESSORIES
Table 600-1. Probe Accessories Specifications (cont)
83RF
HIGH
FREQUENCY PROBE
AC-to-DC Ratio
........................
1:1
Ratio Accuracy (at 1
MHz
and loaded
with 10
MO)
ABOVE
1V
. . . . . . . . . . . . . . . . . . . . . . . . . . ±1 dB
BELOW
1V
..........................
±1.5
dB
Frequency Response (Relative to 1
MHz)
100 kHz to 100
MHz±
1 dB
Extended Frequency Response
.........
Useful for relative readings from 20 kHz
to
250 MHz
Response
.............................
Responds
to
the peak value
of
an
input and is calibrated
to
read rms
value
of
a sine wave
Voltage Range . . . . . . . . . . . . . . . . . . . . . . . . .
0.25
to
30V
rms
Maximum Input Voltage . . . . . . . . . . . . . . . .
30V
rms, 200V de
Input Capacitance . . . . . . . . . . . . . . . . . . . . . Approximately 3 pF
Temperature Range
OPERATING
........................
+10°C
to
+35°C
STORAGE
...........................
-40°C
to
+75°C
Humidity
..............................
<90% R.H.
Output Connector
.....................
Fits standard 0.75-inch dual banana connectors
Accessory
.............................
BNC
to
Probe Adapter, P/N 574756
85RF
HIGH
FREQUENCY PROBE
AC-to-DC Ratio . . . . . . . . . . . . . . . . . . . . . . . .
1:
1
Ratio Accuracy (At 1
MHz
and loaded
with
10
MO)
ABOVE
0.5V
. . . . . . . . . . . . . . . . . . . . . . . . . ±0.5 dB
BELOW 0.5V . . . . . . . . . . . . . . . . . . . . . . . . ±1.0 dB
Frequency Response (Relative to 1
MHz)
100 kHz to 100 MHz
±.5
dB
*100 MHz to 200 MHz ±1.0 dB
*200 MHz to 500 MHz ±3.0 dB
*Referred to high and low inputs at probe tip.
Extended Frequency Response
.........
Useful for relative readings from
20
kHz to 700 MHz
Response
.............................
Responds to the peak value
of
an
input and is calibrated to read rms
value
of
a sine wave
Voltage Range . . . . . . . . . . . . . . . . . . . . . . . . .
0.25
to
30V
rms
Maximum Input Voltage . . . . . . . . . . . . . . . .
30V
rms,
200V
de
Input Capacitance
.....................
Approximately 3 pF
Temperature Range
OPERATING
........................
+10°C
to
+35°C
STORAGE
...........................
-40°C to
+75°C
Humidity
..............................
<90%
R.H.
Output Connector
.....................
Fits standard 0.75-inch dual banana connectors
Accessory
.............................
BNC to Probe Adaoter, P/N 574756
Y8100
DC/AC
CURRENT
PROBE
Ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
20A
ac
or
de
200A
ac
or de
Rated Output
..........................
2V
at
full range
Accuracy
DC
TO
200
HZ
. . . . . . . . . . . . . . . . . . . . . . ±2% of range
200
HZ TO 1 KHZ
...................
<100A add
±3%
reading
>1
OOA
add ±6% reading
Calibration Cycle . . . . . . . . . . . . . . . . . . . . . . 1 year
Frequency Response . . . . . . . . . . . . . . . . . . . de to 1.0 kHz
Recommended Load . . . . . . . . . . . . . . . . . . .
~.O
kO
Temperature Range
....................
+15°C
to +35°C; for specified accuracy
-10°c
to +50°C; storage
and operation at reduced accuracy.
600-7