Table 6-5. Resistance Adjustment Points and Sequence
8508A Range Modes Point Resistance
2 Normal, Normal LoI,
Tru, Tru LoI
Zero 0
Range Gain 1
20 Normal, Normal LoI,
Tru, Tru LoI
Zero 0
Range Gain 10
200 Normal, Normal LoI,
Tru, Tru LoI
Zero 0
Range Gain 100
2 k Normal, Normal LoI,
Tru, Tru LoI
Zero 0
Range Gain 1 k
20 k Normal, Normal LoI,
Tru, Tru LoI
Zero 0
Range Gain 10 k
200 k Normal, Normal LoI Zero 0
Range Gain 100 k
2 M Normal, Normal LoI Zero 0
Range Gain 1 M
20 M Normal, Normal LoI,
HiV
Zero 0
Range Gain 10 M
200 M Normal, Normal LoI,
HiV
Zero 0
Range Gain 100 M
2 G Normal, Normal LoI,
HiV
Zero 0
Range Gain 1 G
20 G HiV Zero 0
Range Gain Optional
1
1. 20 G range gain is adjusted automatically during 2 G HiV range gain adjustment, but may be adjusted
separately (after 2 G HiV range adjustment) at 10 G if a standard is available.
6-20