EasyManua.ls Logo

Fluke 8508A - Page 169

Fluke 8508A
212 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Adjustment and Verification
Routine Adjustment
6
Initial Setup and Connections
1. On the 8508A, select
CLEAR, Pwr Up Dflt to select DCV 1kV range.
2. Ensure that the calibrator output is set to
STANDBY and configured for Internal
Guard (
EX GRD is deselected).
3. Connect the calibrator to the 8508A as shown above in Figure 6-1.
Note
After making the connections allow at least 30 minutes for thermal emfs in
the leads to stabilise.
4. On the calibrator select zero and
OPR to operate the output.
Select
EX SNS.
5. In calibration mode the resolution defaults to
RESL7 in the resistance function. It is
recommended that this resolution is used for adjustment of all resistance ranges.
HiV is restricted to RESL6.
6. Confirm the 8508A is configured for internal guard (External Guard is deselected) by
checking that the ExGd legend does not appear on the left hand display.
Normal
Mode Adjustment
1. Press the
CAL key. Press the key and select 2 range.
2. Press the
CONFIG key and deselect Filt. Ensure LoI is deselected. Ensure 4 W is
selected.
Range Zero and Gain Adjustment Procedure
After the initial setup and connecting up, use the following procedure to adjust range zero
and range gain for the ranges and points detailed in Table 6-5 for the Normal mode.
For the 2 Grange use the 1 G standard resistor in place of the calibrator. Just one
range can be adjusted if required, but for a full-adjustment start with the 2 range and
work up to the 2 G range.
Note that EX SNS (4-wire) is not available on the 5720A 100 M range, and the 8508-
7000K 1 Gstandard resistor is used as a 2-wire device with the calibrator 2-wire
resistance zero used for the corresponding zero points. Ensure
4Wis deselected on the
8508A when adjusting the 200 M and 2 G range, range zero, and range gain points.
On each range, the 8508A automatically recognizes the appropriate value as range zero or
range gain from the applied resistance value.
Note
It is unlikely that the resistance standard will be at exactly the nominal value
required. The Set feature allows a user to enter the true output value of the
calibration standard where it differs from nominal range gain point. If the
resistance value is exactly the nominal value required, omit step 6 for the
Range Gain Point.
6-21

Table of Contents

Other manuals for Fluke 8508A

Related product manuals