10. On the Analyzer, close the TFTP server dialog to terminate the TFTP
session.
11. You are now ready to Install the Software on the Analyzer.
Mapping a Drive and Transferring the Software File from your PC to the Analyzer
Microsoft Networking allows you to share a folder on your PC as a network drive
that you can then access across your local network from your Analyzer, thus
allowing you to copy the software file from your PC to the Analyzer.
Note
You may need to first go to the Network Properties on your Analyzer
(right-click on the Desktop Network Neighborhood and select
Properties), highlight Clients for Microsoft Networking and select
Properties, and setup Logon Validation for your network domain.
Then you will need to restart the Analyzer and log into your network
domain.
The file sharing procedure is slightly different for each Windows operating
system. Generally, you will need to perform the following procedure on your PC:
1. On your PC, determine the directory to be shared (default is C:\temp) that
contains the software file OPVWNA-V2.EXE. From Windows Explorer, set
the folder to be shared by right-clicking on the folder name and selecting
the Sharing tab. Select the Shared As: radio button and enter a Share
Name.
Note
You don’t actually have to copy the software file to the PC hard disk
before transferring it to the Analyzer. You can map the CD-ROM
drive and copy the file directly from the OptiView Wireless Network
Analyzer CD to the Analyzer.
2. Connect the Analyzer to the same local network as the PC.
3. On the Analyzer, open Windows Explorer.
4. Select Tools | Map Network Drive.
5. In the Path field, enter the IP address of the PC followed by the Share
Name (e.g. \\010.010.010.012\sharename .) Click OK when done.
6. Copy the file OPVWNA-V2.EXE from your PC to the Analyzer
(recommended is d:\temp.)
7. You are now ready to Install the Software on the Analyzer.