EasyManua.ls Logo

Fluke QED 6 - Automatic Test Sequence Management

Fluke QED 6
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
QED 6
Users Guide
2-26
Programming an Automatic Test Sequence
The Model H can store in memory up to 28 automatic sequences to fully test
defibrillator performance according to protocol. Standard defaults for
Programs 0-27 for the Model H are listed in Table 2-3.
Table 2-3. Autosequencing Defaults (Model H)
Performance Parameter Default
Print Heading Yes
Energy Measurements 10 J
100 J
200 J
300 J
360 J
Energy Limits + / - 5%
Charge Time Yes
Sync Time Yes
Peak No
Overshoot No
ECG performance 30 BPM
120 BPM
240 BPM
2 Hz Sq
.125 Hz Sq
10 Hz Sin
40 Hz Sin
50 Hz Sin
60 Hz Sin
100 Hz Sin
2 Hz Tri
60 BPM PLS
Fib
Pacer No

Table of Contents

Related product manuals