QED 6
Users Guide
2-34
Fluke Biomedical
Control #:
Serial #:
Model #:
Mfr.
Location:
Technician:
Date:
PROGRAM NAME: PROG 5
Setting Actual Limit +/- 5%
10 J 10.3 J
100 J 10.3 J
200 J 199.0 J
300 J 300.0 J
360 J 360.0 J
Charge Time: 5 sec 198.3 J
Qwave Sync Time 21.0
Rwave Sync Time 50.0
Peak Voltage: 2270 V
Peak Current: 45 A
OVER Voltage: 2 V
OVER Current: 2 A
ECG/Performance Waves
30 BPM
120 BPM
240 BPM
2 Hz Sq
.125 Hz Sq
10 Hz Sin
40 Hz Sin
50 Hz Sin
60 Hz Sin
100 Hz Sin
2 Hz Tri
60 BPM Pls
Vfib
fcf051.eps
Figure 2-11. Automatic Sequence Output with Header