Specifications
Dual Input Oscilloscope
9
97
FLUKE 192B:
10 ns to 2 µs /div.................................up to 500 MS/s
5 µs to 120 s/div .............................................20 MS/s
Record Length
Scope Record Mode ......... >27000 points on each input
Scope Normal Mode .............3000 points on each input
Scope Glitch Capture Mode..............300 min/max pairs
on each input
Glitch Detection
5 µs to 120 s/div......... displays glitches as fast as 50 ns
Waveform Display.................A, B, A+B, A-B, A*B, A vs B
Normal, Average (2,4,8,64 x), Persistence
Time Base Accuracy...................... ±(100 ppm + 0.04 div)
Trigger and Delay
Trigger Modes........................................Automatic, Edge,
External, Video, Pulse Width
N-Cycle (C versions)
Trigger Delay ................................. up to +1200 divisions
Pre Trigger View ............................. one full screen length
Max. Delay......................................................12 seconds
Automatic Connect-and-View Trigger
Source .............................................................. A, B, EXT
Slope.......................Positive, Negative, Dual (C versions)
Edge Trigger
Screen Update ........... Free Run, On Trigger, Single Shot
Source...............................................................A, B, EXT
Slope...................... Positive, Negative, Dual (C versions)
Trigger Level Control Range .......................... ±4 divisions
Trigger Sensitivity A and B
DC to 5 MHz at >5 mV/div.......................... 0.5 divisions
DC to 5 MHz at 2 mV/div and 5 mV/div ...........1 division
200 MHz (FLUKE 199B/C) ..............................1 division
250 MHz (FLUKE 199B/C) ............................ 2 divisions
100 MHz (FLUKE 196B/C) ..............................1 division
150 MHz (FLUKE 196B/C) ............................ 2 divisions
60 MHz (FLUKE 192B)....................................1 division
100 MHz (FLUKE 192B)................................2 divisions
Isolated External Trigger
Bandwidth ............................................................. 10 kHz
Modes .................................................... Automatic, Edge
Trigger Levels (DC to 10 kHz).................... 120 mV, 1.2 V
Video Trigger
Standards..............................PAL, PAL+, NTSC, SECAM
Modes ......................Lines, Line Select, Field 1 or Field 2
Source............................................................................ A
Polarity ................................................. Positive, Negative
Sensitivity ....................................... 0.7 division sync level