Replacing the NiMH Battery Pack BP190
Usually it should not be necessary to replace the battery
pack. However, if replacement is needed, this should be
done by qualified personnel only. Contact your nearest
Fluke center for more information.
Calibrating the Voltage Probes
To meet full user specifications, you need to adjust the red
and gray voltage probes for optimal response. The
calibration consists of a high frequency adjustment and a
dc calibration for 10:1 probes. The dc calibration is not
possible for 100:1 probes.
This example shows how to calibrate the 10:1 voltage
probes:
1
Display the input A key labels.
2
Open the Probe on A menu.
3
Select Voltage, then jump to
Attenuation.
Figure 48. Adjusting Voltage Probes
If the 10:1 option is already selected, proceed with step 5.
4
Select 10:1, then return.