Fluke 192B - 196B/C - 199B/C
Users Manual
110
Electromagnetic Immunity
The Fluke 190 series, including standard accessories, conforms with the EEC directive 89/336 for EMC immunity, as defined
by EN-61326-1, with the addition of the following tables.
Scope Mode (10 ms/div): Trace disturbance with VPS200 voltage probe shorted
Table 1
No visible disturbance E = 3V/m
Frequency range 10 kHz to 20 MHz 2 mV/div to 100 V/div
Frequency range 20 MHz to 100 MHz 200 mV/div to 100 V/div
Frequency range 100 MHz to 1 GHz 500 mV/div to 100 V/div
*)
(*) With the 20 MHz Bandwidth Filter switched on: no visible disturbance.
With the 20 MHz Bandwidth Filter switched off: disturbance is max 2 div.
Table 2
Disturbance less than 10% of full scale E = 3V/m
Frequency range 20 MHz to 100 MHz 10 mV/div to 100 mV/div
Test Tool ranges not specified in tables 1 and 2 may have a disturbance of more than 10% of full scale.
Meter Mode (Vdc, Vac, Vac+dc, Ohm and Continuity): Reading disturbance with test leads shorted
Table 3
Disturbance less than 1% of full scale E = 3V/m
Frequency range 10 kHz to 1 GHz 500mV to 1000V , 500Ohm to 30 MOhm ranges