Specifications
10:1 Probe
9
109
10:1 Probe
Safety
Max. Input Voltage............................ 1000 V CAT II
600 V CAT III
Max. Floating Voltage
from any terminal to earth ground ........... 1000 V CAT II
600 V CAT III
Electrical specifications
Input Impedance at probe tip ......10 MΩ ±2 %//14 pF ±2 pF
Capacity Adjustment Range............................10 to 22 pF
Attenuation at DC (1 MΩ input) .................................. 10 x
Bandwidth (with FLUKE 199C) .....DC to 200 MHz (-3 dB)
Accuracy
Probe accuracy when adjusted on the test tool:
DC to 20 kHz..........................................................±1 %
20 kHz to 1 MHz ....................................................±2 %
1 MHz to 25 MHz ...................................................±3 %
For higher frequencies the probe’s roll off starts affecting
the accuracy
Environmental
Temperature
Operating..............................0 to 50 °C (32 to 122 °F)
Storage.......................... -20 to +60 °C (-4 to +140 °F)
Altitude
Operating........................................3 km (10 000 feet)
Storage.........................................12 km (40 000 feet)
Humidity
Operating at 10 to 30 °C (50 to 86 °F) ............... 95 %
0.01
1
2
5
10
20
50
100
200
500
1000
0.02 0.05 0.1 0.2 0.5 1 2 5 10 20 50 100 20
FREQUENCY (MHz)
.
VOLTAGE (Vrms)
CAT III
CAT II
Figure 51. Max. Voltage From Probe Tip to earth
ground and From Probe Tip to Probe Reference