Fluke 192B - 196B/C - 199B/C
Users Manual
102
Meter
Meter Input
Input Coupling..............................................................DC
Frequency Response.......................DC to 10 kHz (-3 dB)
Input Impedance ................ 1 M (±1 %)//10 pF (±1.5 pF)
Max. Input Voltage ..............................1000 V CAT II
600 V CAT III
(For detailed specifications, see “Safety”)
Meter Functions
Ranging ....................................................... Auto, Manual
Modes.....................................................Normal, Relative
DMM Measurements on Meter Inputs
The accuracy of all measurements is within ± (% of
reading + number of counts) from 18 °C to 28 °C.
Add 0.1x (specific accuracy) for each °C below 18 °C or
above 28 °C.
General
DC Common Mode Rejection (CMRR) ..................>100 dB
AC Common Mode Rejection at 50, 60, or 400 Hz ......>60 dB
Ohms (
)
Ranges..................................500.0 , 5.000 k, 50.00 k,
500.0 k, 5.000 M, 30.00 M
Full Scale Reading
500 to 5 M...............................................5000 counts
30 M ...........................................................3000 counts
Accuracy ............................................... ±(0.6 % +5 counts)
Measurement Current ................... 0.5 mA to 50 nA, ±20 %
decreases with increasing ranges
Open Circuit Voltage ....................................................<4 V
Continuity (CONT)
Beep.......................................................... <50 (±30 )
Measurement Current ...............................0.5 mA, ±20 %
Detection of shorts of ............................................. 1 ms