Specifications
Automatic Scope Measurements
9
101
Power
Power Factor........................ ratio between Watts and VA
Range .........................................................0.00 to 1.00
Watt ................................... RMS reading of multiplication
corresponding samples of input A (volts)
and Input B (amperes)
Full Scale Reading ....................................... 999 counts
VA................................................................ Vrms x Arms
Full Scale Reading ....................................... 999 counts
VA Reactive ..................................................
((VA)
2
-W
2
)
Full Scale Reading ....................................... 999 counts
Phase
Range ............................................ -180 to +180 degrees
Resolution.......................................................... 1 degree
Accuracy
0.1 Hz to 1 MHz ...........................................±2 degrees
1 MHz to 10 MHz .........................................±3 degrees
Temperature (TEMP)
With Optional Temperature Probe
Ranges (°C or °F)...................................-40.0 to +100.0 °
-100 to +250 °
-100 to +500 °
-100 to +1000 °
-100 to + 2500 °
Probe Sensitivity ............................1 mV/°C and 1 mV/°F
Decibel (dB)
dBV ................................................dB relative to one volt
dBm....................dB relative to one mW in 50 or 600
dB on.......................................... VDC, VAC, or VAC+DC
Accuracy .......................... same as VDC, VAC, VAC+DC