Fluke 192B - 196B/C - 199B/C
Users Manual
100
Amperes (AMP)
With Optional Current Probe or Current Shunt
Ranges .............................same as VDC, VAC, VAC+DC
Probe Sensitivity ................. 100 µV/A, 1 mV/A, 10 mV/A,
100 mV/A, 1 V/A, 10 V/A, and 100 V/A
Accuracy...........................same as VDC, VAC, VAC+DC
(add current probe or current shunt accuracy)
Peak
Modes............................Max peak, Min peak, or pk-to-pk
Maximum Voltage
with 10:1 probe .................................................. 1000 V
direct (1:1)............................................................ 300 V
Maximum Resolution
with 10:1 probe ................................................... 10 mV
direct (1:1)............................................................. 1 mV
Full Scale Reading.......................................... 800 counts
Accuracy
Max peak or Min peak................................±0.2 division
Peak-to-peak..............................................±0.4 division
Frequency (Hz)
Range ..................................... 1.000 Hz to full bandwidth
Full Scale Reading .......................................9 999 counts
with at least 10 waveform periods on screen.
Accuracy
1 Hz to full bandwidth ...................... ±(0.5 % +2 counts)
Duty Cycle (DUTY)
Range ..................................................... 4.0 % to 98.0 %
Pulse Width (PULSE)
Resolution (with GLITCH off) ........................ 1/100 division
Full Scale Reading ..........................................999 counts
Accuracy
1 Hz to full bandwidth ...................... ±(0.5 % +2 counts)
Vpwm (C versions only)
Purpose................. to measure on pulse width modulated
signals, like motor drive inverter outputs
Principle .........readings show the effective voltage based
on the average value of samples over a
whole number of periods of the fundamental
frequency
Accuracy ............................ as Vrms for sinewave signals