Specifications
Environmental
9
107
Environmental
Environmental......................... MIL-PRF-28800F, Class 2
Temperature
Operating:
battery only........................... 0 to 50 °C (32 to 122 °F)
power adapter ...................... 0 to 40 °C (32 to 104 °F)
Storage .............................-20 to +60 °C (-4 to +140 °F)
Humidity
Operating:
0 to 10 °C (32 to 50 °F) .......................noncondensing
10 to 30 °C (50 to 86 °F) .................................... 95 %
30 to 40 °C (86 to 104 °F) ...................................75 %
40 to 50 °C (104 to 122 °F) .................................45 %
Storage:
-20 to +60 °C (-4 to +140 °F)...............noncondensing
Altitude
Operating ..........................................3 km (10 000 feet)
Storage ........................................... 12 km (40 000 feet)
Vibration (Sinusoidal).......................................... max. 3 g
Shock................................................................ max. 30 g
Electromagnetic Compatibility (EMC)
Emission and immunitiy ............ EN-IEC61326-1 (1997)
Enclosure Protection............................. IP51, ref: IEC529
Safety
Designed for 1000 V measurements Category II , 600 V
measurements Category III , Pollution Degree 2, per:
ANSI/ISA S82.01-1994
EN/IEC61010-1 : 2001
CAN/CSA-C22.2 No.1010.1-92
UL3111-1
Max. Input Voltages
Input A and B directly ............................... 300 V CAT III
Input A and B via 10:1 probe ................... 1000 V CAT II
600 V CAT III
METER/EXT TRIG inputs........................ 1000 V CAT II
600 V CAT III
Max. Floating Voltage
From any terminal to earth ground .......... 1000 V CAT II
600 V CAT III
Between any terminal .............................. 1000 V CAT II
600 V CAT III
Voltage ratings are given as “working voltage”. They
should be read as Vac-rms (50-60 Hz) for AC sinewave
applications and as Vdc for DC applications.