Using The Recorder Functions
Recording Scope Waveforms In Deep Memory (Scope Record)
3
39
Using External Triggering to Start or Stop Scope
Record
To record an electrical event that causes a fault, it might
be useful to start or stop recording on an external trigger
signal:
Start on trigger to start recording; recording stops when
the deep memory is full
Stop on trigger to stop recording.
Stop when untriggered to continue recording as long as
a next trigger comes within 1 division in view all mode.
To set up the test tool, continue from step 3 of the
previous section:
4 Apply the signal to be recorded to the red BNC
input A. Apply a trigger signal to the red and
black external trigger banana inputs. (See Figure
20.)
5
Open the Recorder Options
menu.
Figure 20. Scope Record Using External Triggering
6
Jump to Display Glitches:.
7
Jump to Mode:.
8
Select on EXT. ... to open the
Single Sweep on Ext. menu.