EasyManuals Logo

Fluke ScopeMeter 192C Service Manual

Fluke ScopeMeter 192C
86 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #56 background imageLoading...
Page #56 background image
Fluke 19xB-19xC-2x5C
Service Manual
4-32
4.9 Probe Calibration Generator Test
To verify the internal probe calibration square wave generator, you can do a Probe
Calibration as described in section 5.8. If no square wave appears on the screen, either
the probe is defective: try another probe, check the probe with an external voltage in a
scope application,
or
the internal square wave generator is defective.
This is the end of the Performance Verification Procedure.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke ScopeMeter 192C and is the answer not in the manual?

Fluke ScopeMeter 192C Specifications

General IconGeneral
BrandFluke
ModelScopeMeter 192C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals