Calibration Adjustment
5.6 Final Calibration 5
5-15
al55ex2w.bmp
Figure 5-7. 5500A NORMAL Output to Test Tool Banana Input
3. Set the 5500A to supply a DC voltage, to the first calibration point in Table 5-5.
4. Set the 5500A to operate (OPR).
5. Press
to start the calibration.
6. Wait until the display shows calibration status
:READY.
7. Press
to select the next calibration step, set the 5500A to the next calibration
point, and start the calibration. Continue through all calibration points of Table 5-4
8. Set the 5500A to Standby, and continue at Section 5.6.7.
Table 5-5. DMM Gain Calibration Points
Cal step UUT input value (5500A NORMAL)
CL 0840 500 mV
CL 0849 2.5 V
CL 0841 5 V
CL 0842 50 V (set 5500A to OPR!)
CL 0843 500 V
CL 0844 1000 V
5.6.7 Input A& B, and DMM Zero
Proceed as follows to do the Input A&B, and the DMM Zero calibration:
1. Press
to select calibration adjustment step CL0852
2. Short circuit Input A and Input B.