Ti200, Ti300, Ti400, Ti450, Ti450SF6, Ti480
Users Manual
34
Table 17 lists the guidelines for the Imager to detect a gas leak based on the annual gas loss rate.
SmartView Software
SmartView software for a PC is available to use with the Imager and contains features to analyze
images, organize data and information, and make professional reports.
Use SmartView software to:
• Review IR-PhotoNotes, audio, and text annotations.
• Export IR and visible images.
• Edit .is2 image files and .is3 video files.
• Enable Bluetooth, WiFi, and Fluke Connect functions.
• Update the firmware for new Imager feature
Table 17. Detection Guidelines
Annual Gas
Loss Rate
Guidelines
<10 lbs (<4.5 kg)
• The greatest possible difference between the gas and the background
environment temperatures
• Wind speed: <1.12 mps (<2.5 mph)
• Uniform background scene (clear blue sky)
• Imager on a tripod, close to the gas leak.
10 lbs to 50 lbs
(4.5 kg to 22,7 kg)
• ≥10 °C difference between the gas and the background environment
temperatures
• Wind speed: <2.24 mps (<5 mph)
• Variable background scene (clear blue sky, high altitude clouds okay)
• Imager on a tripod.
>50 lbs (>22,7 kg)
• ≥3 °C difference between the gas and the background environment
temperatures
• Wind speed: <4.47 mps (<10 mph)
• Variable background scene (high altitude clouds okay)
• Imager on a tripod or handheld.