2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 159
NW- 2.0.0 VALVE #41’S ‘DOT ALWAYS BLUE = CONSTANT “NO WATER” ALARM
a) Figure below, ENSURE Valve #41’s connector is plugged PROPERLY into distribution board position
V27, IN-V”.
b) Call debug screen 0 to watch Valve #41’s ’dot’ for one (1) FULL minute. TWO (2) possible scenarios:
1) IF (and ONLY if) Valve #41’s ‘dot’ is now remaining white OR cycling between white and
blue: The constant “No Water” alarm is no longer occurring! If (and ONLY if) a “No Water” alarm
occurs intermittently return to (ABOVE) procedure number NW- 1.0.10 (
page 157). If a “No
Water” alarms NEVER occurs the No Water problem is not occurring!
2) IF Valve #41’s ‘dot’ is still NEVER white i.e. ALWAYS blue:
Ignoring Flow Error, look at the
2
nd
window down, Valve
Error! TWO (2) possible scenarios:
Scenario #1: IF (and ONLY if) Valve Error = 0! See procedure number NW- 2.0.1 (page 160).
Scenario #2: IF Valve Error = 1 longer than two (2) seconds consistently! Refer to the
Table below:
If the machine is CURRENTLY in:
Your response
Dialysis Program i.e. connected to
acid and bicarbonate
Proceed to
page 711, Section 26
A Cleaning / Disinfection Program
(Rinse, Heat Disinfect, etc.)
Proceed to
page 207, procedure number CLEAN- 7.0.0

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals