EasyManuals Logo

Fresenius Medical Care 2008T Troubleshoot Manual

Fresenius Medical Care 2008T
768 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #388 background imageLoading...
Page #388 background image
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 384
CO- 6.2.0 CALIBRATE / TROUBLESHOOT COND CELLS
a) Install the hydraulics but do not screw in the rear panels.
b) ENSURING plenty of known good acid and LIQUID Bicarb plug FIRMLY into each.
c) Place the machine into Service Mode Calibrate Sensors
Cond Cells (if equipped with a bibag Connector) Dialysate
Cells. DO NOT select Bicarb Cell!
d) Install a Temperature / Conductivity meter into the dialysate
lines. Figure right, flow through the meter MUST directed from
bottom to top.
e)
Close the shunt door completely!
f) Follow the screen’s instructions through: 4.
Waiting until the Conductivity value is
stable.” The screen’s [Conductivity] data
box is initially gray.
g) ENSURING the external flow indicator’s ‘bobis moving at least ¼ way up the sight tube ALLOW six
(6) minutes before continuing to STEP #1 below:
STEP #1: Set the external meter to measure Temperature (°C). Watch it for thirty (30) seconds.
Between 35.1 and 39.9° C AND stable i.e. DOES NOT change more than +/- 0.25° C?
Yes Temperature between 35.1 and 39.9° C! Continue to STEP #2.
No a) Return to DIALYSIS PROGRAM (“Select Program” → ‘Dialysis’ → ‘Enter)!
b) Proceed to page 232, SECTION 4 TEMPERATURE PROBLEMS
STEP #2: Set the meter to measure Conductivity (mS). Its reading MUST BE between 13.0 and 14.5
mS?
Yes Between 13.0 and 14.5 mS! Continue to STEP #3 next page.
No NOT between 13.0 and 14.5 mS! See parts a AND b below.
a) SLAM the Loading Pressure gauge into the Acetate/Acid rinse port. If no masked
“No Water” or Flow Error alarms the gauge cycles, approximately EVERY three (3)
seconds, to about 25 psi. If not cycling return to Dialysis Program!
b) Proceed to
page 391, procedure number CO- 6.8.0.
NOTE! If an “Operator Error” banner appears but the flow indicator’s ‘bob continues to
move up and down ignore the banner. If the ‘bob’ NOT moving ‘Escape’ calibration
THEN return to procedure number CO- 6.2.0 (page 384).
While in Dialysis, ENSURE the machine is clear of “No Water”, Flow Errors and
Temperature alarms. Hydraulic alarm banners are not displayed in Service mode!
These are TROUBLESHOOTING procedure! Perform EXACTLY as written to avoid error!

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals