2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 526
PHT- 3.1.5 ALL OTHER SCENARIOS / PRESSURE TEST FAILED
Has the DiaSafe
®
filter been replaced in THIS Troubleshooting session?
Yes The DiaSafe
®
filter has been replaced! Proceed to page 483, SECTION 9 TMP
PROBLEMS
No a) Replace the DiaSafe
®
filter preferably with one from another machine that is passing
Pressure Tests.
b) Place the machine into RINSE for five (5) minutes!
c) Return to Dialysis Program (“Select Program” → ‘Dialysis’ → ‘Enter’).
d) From the Home screen, set [Dialysate Flow] to 800 ml/min and press ‘Enter’!
e) Call debug screen 0 and allow Valve #24’s ‘dot’ to stay blue!
f) Allow two (2) minutes BEFORE continuing to part g.
g) Reset all alarms and repeat the Pressure Tests. Do they BOTH pass?
Yes Both Pass! The DiaSafe
®
filter MAY have been causing the problem
HOWEVER, see procedure number PHT- 4.0.0 (page 527).
No Pressure test(s) Fail! Proceed to page 483, SECTION 9 TMP PROBLEMS
LEFT BLANK INTENTIONALLY

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals