2008T Troubleshooting Guide
P/N 490292 Rev. A
PHT- 3.1.5 ALL OTHER SCENARIOS / PRESSURE TEST FAILED
Has the DiaSafe
®
filter been replaced in THIS Troubleshooting session?
Yes The DiaSafe
®
filter has been replaced! Proceed to page 483, SECTION 9 – TMP
PROBLEMS
No a) Replace the DiaSafe
®
filter preferably with one from another machine that is passing
Pressure Tests.
b) Place the machine into RINSE for five (5) minutes!
c) Return to Dialysis Program (“Select Program” → ‘Dialysis’ → ‘Enter’).
d) From the Home screen, set [Dialysate Flow] to 800 ml/min and press ‘Enter’!
e) Call debug screen 0 and allow Valve #24’s ‘dot’ to stay blue!
f) Allow two (2) minutes BEFORE continuing to part g.
g) Reset all alarms and repeat the Pressure Tests. Do they BOTH pass?
Yes Both Pass! The DiaSafe
®
filter MAY have been causing the problem
HOWEVER, see procedure number PHT- 4.0.0 (page 527).
No Pressure test(s) Fail! Proceed to page 483, SECTION 9 – TMP PROBLEMS
LEFT BLANK INTENTIONALLY