EasyManuals Logo

Fresenius Medical Care 2008T Troubleshoot Manual

Fresenius Medical Care 2008T
768 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #406 background imageLoading...
Page #406 background image
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 402
CO- 8.2.3 BIBAG EQUIPPED.
Call debug screen 14. WITHOUT LOOKING AWAY watch Air (4
th
column from left, 3
rd
window down) for one (1) minute. Does Air ever = 1?
3) IF (and ONLY if) Air = 0 ALWAYS (NEVER = 1): See procedure number CO- 8.2.4 (page 403).
4) IF Air = 1: See parts a THROUGH e below:
a) ENSURE no air through the clear Bicarbonate inlet tubing! This may cause Air = 1!
b) ENSURE screen 14’s Conductivity (2
nd
column from left, bottom row) is MORE THAN 48.00
indicating the bicarb AND Bicarb Pump is good. Bad bicarbonate will cause AIR = 1!
c) Remove the hydraulics from the cabinet!
d) Figure below, ENSURE the blue and brown wires are attached securely to their probes on top of
the bibag Air Separator Chamber! If not this will cause AIR = 1!
e) If Air still = 1 FOUR (4) possible problems: 1) Loose bibag Interface Board ribbon cable (refer to
Figure 4C (page 11)) OR; 2) Bad bibag Interface Board OR; 3) Bad connection, from Air Sensor
#112, inside the bibag Distribution Board (refer to Figure 4D (page 13) OR; 4) Bad bibag Air
Separator Chamber.

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals