223
5.2Maintenance
DX100 Service Manual
5.2.6 IES Detection Flag
1. Select 1. Adjustment / Maintenance ï‚® Maintenanceï€ ï‚® IES Detection Flag.
2. S
elect [Before repairing] and click [Run].
Make sure that the status of the [Current] changed to [Under repair].
3. After replacing Ink Cartridge Holder, make sure that [Normal] is displayed in the
status of the [Current].
Figure 5-10. IES Detection Flag Screen
5.2.7 APG Check
1. Select 3. Function Check ï‚® Inspectionï€ ï‚® APG Check.
2. Click [RUN].
Figure 5-11. APG Check Screen
3. If judged as NG, re-install the APG Unit and perform the APG Check again.