ActionStep
Connect a 25 to 30 mL syringe to the purge valve of one of the pump heads.4
Open the purge valve by turning it counter-clockwise one and a half turns.
Draw liquid slowly into the syringe until liquid reaches the pump and the inlet
tubing is filled with liquid.
5
Close the purge valve by turning it clockwise. Disconnect the syringe and
discard its contents.
6
Repeat steps 3 to 6 for each inlet that is to be used during the run.7
ÄKTA go Operating Instructions 29360951 AC 65
5 Prepare the system for a run
5.2 Prime inlets and purge pump heads