ActionStep
Clean the conductivity flow cell before calibration (see Section 5.3.1 Clean
the conductivity flow cell, on page 96).
1
Disconnect the tubing from the bottom of the conductivity monitor and re-
place it with a Luer-to-SNAP connector.
2
Disconnect the tubing from the top of the conductivity monitor and replace
it with a piece of tubing leading to the waste container.
3
W
Select System:Settings in UNICORN System Control module.4
In the Instructions list, select Conductivity:Cond Temp Compensation.5
Set the Compensation factor to 0 and click OK.6
Select System:Calibrate from the main System control menu.7
ÄKTA pilot 600 User Manual 29274559 AA 105
5 Maintenance
5.4 Calibration procedures
5.4.2 Calibrate the conductivity monitor