ActionStep
Select whether the test report should be saved to a file and/or printed.
Note:
Sending the report to a printer requires that a system printer is installed on
the UNICORN Instrument Server (see the UNICORN Administration and Tech-
nical Manual for details).
6
Click Next> to continue.
Specify the details of the result file from the performance test.
Note:
The result file is separate from the test report. The report will be generated
even if No result is selected.
7
Click Start to start the performance test.
Follow any instructions that are shown on the screen.8
The progress of the performance test is shown in the chromatogram pane and can be
customized in the same way as any chromatogram.
Required materials
Materials required for performance tests, in addition to distilled water, are specified in
the method notes for the test and summarized in the table below.
Required materialTest
2 L 1% acetone (v/v) in 1.0 M NaClSystem test/System test pre
column cond
Reference capillary 1
Reference capillary 2Column valve test
Short lengths of i.d. 3.2 mm with SNAP connectors
to replace columns.
Reference capillary 2Pressure pulsation test
124 ÄKTA pilot 600 User Manual 29274559 AA
6 Performance tests
6.2 How to run performance tests