Page 122 Section 8.0 - Storage Requirements
Section 8.0
Storage Requirements
8.1 Short Term Storage (Less than 6 months)
If the CT system is to be stored before installation, store in a temperature and humidity controlled
warehouse. Protect from weather, dirt and dust.
Meeting these requirements prevents rust and corrosion from forming on bearing surfaces due to
condensation.
Storage temperature should not exceed 40º to +80º F (4º to +27º C).
Maintain relative humidity (non-condensing) between 20% and 60%.
Maximum relative humidity rate of change is 5%/hr.
The maximum temperature rate of change is 5
o
F/hr. (3
o
C/hr.)
Air pressure should be between 700hPa and 1060hPa.
NOTICE Between delivery is considered short-term storage.
Van storage must meet the same specifications as above.
8.2 Construction-Site Storage
When storing the CT system at a construction site be sure to adhere to the following storage
requirements:
Do not damage or puncture the shipping crate.
Do not remove packaging until all construction is completed at the site and all dust created by
the construction is removed.
Maintain a storage temperature within the range of 10º to 32º C (50º to 90º F).
Maintain a relative humidity (non-condensing) between 20% and 70%.
8.3 Long Term Storage (6 months or more)
If the CT system is to be stored before installation, store in a temperature and humidity controlled
warehouse. Protect from weather, dirt and dust.
Meeting these requirements prevents rust and corrosion from forming on bearing surfaces due to
condensation.
Storage temperature should not exceed 50º to +90º F (10º to +32º C).
Maintain relative humidity (non-condensing) between 20% and 70%.
If you are storing a system longer than six (6) months, but less than 12 months, contact CT
Engineering for installation start-up instructions. Storage longer than 12 months is not
recommended.