IRECTION 5141177-100, REVISION 14 BRIGHTSPEED ELITE, EDGE, EXCEL: PRE-INSTALLATION
Page 86 Section 8.0 - Structural Requirements
8.5.3 Equipment Location
To minimize the interference, the CT equipment should be placed on a solid floor, located as far as
possible from the following vibration sources:
8.6 Walls: Scan Window
The recommended patient viewing window dimensions are 1219 mm wide by 1067 mm high (48 in.
x 42 in. ). The location of the window is dependent on the position of Operator Workspace position.
Consult Section 10.0 of this chapter and a qualified radiological health physicist for radiation
protection requirements of glass (lead content and thickness).
Note: The operator at the Operator Workspace must be able to view the patient during a scan.
8.7 Ceiling Requirement
The weight of "Boom in a room LCD monitor" assy is 50 kg., and ceiling shall have enough strength
against the specification.
• Parking lots • Roadways • Subways
• Trains • Hallways • Elevators
• Heliports • Hospital power plants containing pumps, motors, air handling
equipment and air conditioning units