GE HEALTHCARE
D
IRECTION 5141177-100, REVISION 14 BRIGHTSPEED ELITE, EDGE, EXCEL: PRE-INSTALLATION
Page 145 Section 3.0 - Interconnect Runs, Wiring and Cables
Table 9-10 GE Healthcare Supplied Cables for NIO16 Console (Optional, Long Run) (2281840-13)- UL Information
Note: For long cable kits GEHW Cat# is B70992RT, WSO Cat# is B7580JZ
RUN #
LENGTH,
ACTUAL
(USABLE)
UL CABLE INFORMATION
UL Style
Flam.
Rating
Voltage
Rating
Voltage
Actual
Temp.
Rating (C)
Dia.
mm (inch)
# of Cond
Size AWG
PULL SIZE
MM (INCHES)
ft m
PART # DESCRIPTION
050 63
(55)
19.3
(16.76)
2343529 HVDC, PDU to Gantry 2587 FT4 600 + & - 350VDC 90 19 (.751) 3 (2) 4
(1) 8
22 (.87) Dia
051 62.5
(55)
19
(16.76)
2343530 HVAC, PDU to Gantry 2587 FT4 600 440Y/254 90 15.3 (.604) 4 14 11.2 (.44) Dia
052 63
(58)
19.3
(17.56)
2343528 LVAC, PDU to Gantry 2587 FT4 600 208Y/120 90 13.8 (.542) 5 8 56.4 (2.22) Dia
053 80
(75)
24.5
(22.86)
2343531 LVAC, PDU to Console 2587 FT4 600 120VAC 90 12.3 (.483) 3 10 56.4 (2.22) Dia
055 63
(55)
19.3
(16.76)
2371450 Ground, PDU to Raceway 1284 VW-1
(FT-1)
600 0 105 15.5 (.608) 1 1/0 15.8 (.62) Dia
056 83
(75)
25.5
(22.86)
2371450-3 Ground, Raceway to Console 1283 VW-1
(FT-1)
600 0 105 11.9 (.467) 1 2 12.2 (.48) Dia
100 70
(62)
21.4
(18.86)
5120646 Signal, Gantry MSUB or
TGPU to PDU
FT-4 300 <30VDC 80 11.2 (.440) 25 22 17 x 58 (.68 x 2.30)
19 x 51 (.75 x 2.01)
101 86
(78)
26.35
(23.71)
5419981 Signal, Gantry MSUB or
TGPU to OC
FT-4 300 <30VDC 80 11.2 (.440) 25 22 17 x 58 (.68 x 2.30)
19 x 51 (.75 x 2.01)
102 86
(81)
26.3
(24.84)
2373436-2 Signal (LAN), Gantry to OC 1900 <30VDC 5.9 (.234) 8 24 15 (.59) Dia
103 80
(75)
24.3
(22.86)
5432019 Fiber Optic, Gantry to OC N/A N/A 1 N/A 10 (.39) Dia