AC.24.078
2. Push the latch handle down to release the cover.
AC.24.079
3. Pull the latch hook away from the canister and lift the cover off
the canister.
AC.24.082
4. Remove and properly discard the absorbent and any water in
the canister body.
Be careful when emptying the absorber canister. The
liquid is caustic and may burn skin.
5. To clean and disinfect the canister, refer to "Automated washer"
in the "Cleaning and sterilization" manual.
6. Pour fresh absorbent into the canister.
7. Gently shake the canister from side to side to level the top of the
absorbent. Make sure the absorbent in the canister does not
exceed the fill line.
8. Wipe off any absorbent dust on the top cover wiper seal and the
canister sealing area.
The top cover wiper seal may discolor over time, however color
change alone has no effect on seal function.