9. Insert the cover into the groove on the canister handle so that
the arrows get covered by the black zone.
AC.24.081
10. Attach the cover latch hook to the canister body. Push the latch
handle down until it is flush with the canister cover.
AC.24.368
Failure to properly lock the canister cover could cause
canister malfunction and breathing circuit leaks. The
locking tab of the canister cover must be in the
downward position and flush with the cover.
11. Place the canister onto the canister lifter base.
12. Pull the canister lifter handle up to lock the canister into the
breathing system.
13. Always perform a "Preoperative checkout" after reassembly
before using the anesthesia system.