Monitoring ECG
2000966-386D Dash™ 3000/4000/5000 8-13
Arrhythmia alarm levels
HR, PVC and ST parameter alarm levels
ECG SIZE
DETECT PACE
ARRHYTHMIA
ST ANALYSIS
LEAD ANALYSIS
HR HIGH LIMIT and HR LOW LIMIT
PVC LIMIT and PVC state
To select the source of ECG settings, follow this procedure.
1. Select MORE MENUS > MONITOR SETUP > MONITOR DEFAULTS >
SETUP DEFAULT DISPLAY.
2. Locate and select COMBO DEFAULT SOURCE.
3. Select one of the following:
TELE DEFINED — Use the telemetry ECG settings. This is the factory
default.
USER DEFINED — Use the monitors ECG settings if telemetry is NOT
admitted. Use the telemetry ECG settings if telemetry IS admitted.
ECG Setting Source When Entering COMBO Mode
The following tables indicate the source (monitor or telemetry) of ECG settings when
entering COMBO mode.
Notice that selecting Tele Defined means the ECG setting source is always telemetry
when entering COMBO mode, while the ECG setting source for User Defined
depends on whether the patient is admitted to the monitor or telemetry when entering
COMBO mode.
Tele Defined
Monitor Telemetry Entering COMBO Mode ECG Setting Source
Not admitted Not admitted >
Telem etryAdmitted Not admitted >
Not admitted Admitted >
User Defined
Monitor Telemetry Entering COMBO Mode ECG Setting Source
Not admitted Not admitted >
Monitor
Admitted Not admitted >
Not admitted Admitted > Telemetry

Table of Contents

Other manuals for GE Dash 4000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 4000 and is the answer not in the manual?

GE Dash 4000 Specifications

General IconGeneral
Resolution640 x 480 pixels
Power Supply100-240 VAC, 50/60 Hz
ECG3/5/12-lead
IBPUp to 4 channels
Battery BackupYes
CertificationsCE, UL, CSA
Parameters MonitoredECG, SpO2, NIBP, IBP, Temperature, Respiration, Cardiac Output
ConnectivityEthernet, USB, Serial
Display TypeTFT LCD
RecorderIntegrated thermal printer
Battery TypeLithium-ion
Operating Temperature0 to 40°C
Storage Temperature-20 to 60°C
Humidity Range15% to 95% non-condensing
Display Size10.4 inches
Cardiac OutputAvailable