Maintenance
2000966-386D Dash™ 3000/4000/5000 A-17
d. Select CHECK CALC IN PROGRESS > STOP to deflate the NBP cuff.
8. Repeat procedure for the BP 2/4 patient connector.
Alarm Verification
To verify the system alarm functions, use the following procedure.
1. Raise the heart rate limit until it is higher that the patient’s current heart rate.
2. Verify the following:
The audible alarm sounds the correct tone.
The heart rate in the ECG parameter window flashes in the correct color.
The alarm light flashes the correct color.
An alarm graph printed.
3. Silence the alarms and verify they react correctly.
4. Return the heart rate alarm settings to the original monitoring levels.
NOTE
For more information on alarms, refer to Chapter 4.

Table of Contents

Other manuals for GE Dash 4000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 4000 and is the answer not in the manual?

GE Dash 4000 Specifications

General IconGeneral
Resolution640 x 480 pixels
Power Supply100-240 VAC, 50/60 Hz
ECG3/5/12-lead
IBPUp to 4 channels
Battery BackupYes
CertificationsCE, UL, CSA
Parameters MonitoredECG, SpO2, NIBP, IBP, Temperature, Respiration, Cardiac Output
ConnectivityEthernet, USB, Serial
Display TypeTFT LCD
RecorderIntegrated thermal printer
Battery TypeLithium-ion
Operating Temperature0 to 40°C
Storage Temperature-20 to 60°C
Humidity Range15% to 95% non-condensing
Display Size10.4 inches
Cardiac OutputAvailable

Related product manuals