EasyManuals Logo

GE Dash 4000 Operator's Manual

GE Dash 4000
270 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #132 background imageLoading...
Page #132 background image
10-8 Dash™ 3000/4000/5000 2000966-386D
Monitoring NBP
CUFF SIZE Select the type of NBP cuff used on the patient (ADULT,
PEDIATRIC or NEONATAL).
WARNING
—The correct cuff size must be selected to obtain reliable
NBP data and to prevent overpressure in neonatal or
pediatric use.
NOTE
You must select CLEAR NBP READINGS after
changing this setting for the target pressure to take
affect.
CLEAR NBP READINGS Remove all values from the NBP parameter window, remove
NBP messages from the display and remove all NBP
information from the VITAL SIGNS history.
SETUP CUSTOM AUTO
CAUTION
—Changes made to these control settings are persistent.
They are recalled when a patient is discharged from the
monitor.
NOTE
Only available in software version 6 or later.
Create a custom series of automatic NBP measurements,
including the time interval between measurements and the
number of measurements to perform. For more information,
refer to Setup Custom Automatic NBP Measurements on
page 10-9.
INITIAL INFLATION
PRESSURE
Select the target inflation pressure for the first NBP
measurement. The choices vary by cuff size.
NOTE
You must select CLEAR NBP READINGS after
changing this setting for the target pressure to take
affect.
NOTE
SuperSTAT NBP briefly over-inflates beyond the initial
inflation pressure and then steps down to the correct
pressure. To benefit patient comfort, lower the default
initial inflation pressure.
PULSE RATE NOTE
Only available in software version 6 or later.
Turn the feature on or off to display pulse rate values in the
parameter window when the display setting is INDV 3 WFS.
Option Function

Table of Contents

Other manuals for GE Dash 4000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 4000 and is the answer not in the manual?

GE Dash 4000 Specifications

General IconGeneral
Resolution640 x 480 pixels
Power Supply100-240 VAC, 50/60 Hz
ECG3/5/12-lead
IBPUp to 4 channels
Battery BackupYes
CertificationsCE, UL, CSA
Parameters MonitoredECG, SpO2, NIBP, IBP, Temperature, Respiration, Cardiac Output
ConnectivityEthernet, USB, Serial
Display TypeTFT LCD
RecorderIntegrated thermal printer
Battery TypeLithium-ion
Operating Temperature0 to 40°C
Storage Temperature-20 to 60°C
Humidity Range15% to 95% non-condensing
Display Size10.4 inches
Cardiac OutputAvailable

Related product manuals