11-6 Dash™ 3000/4000/5000 2000966-386D
Monitoring SpO2
WARNING
DATA VALIDITY—Do not expose sensor detector to strong
ambient light while monitoring a patient. A poor signal may result.
WARNING
DATA VALIDITY —Do not allow tape to block the sensor
photodetectors.
WARNING
PATIENT SAFETY—Prolonged monitoring may require changing
the sensor site periodically. Move the sensor if there is any sign of
skin irritation or impaired circulation. Change the sensor site at least
every four hours to prevent ischemic skin necrosis. Be particularly
careful when monitoring neonates. If required, reduce the
application periods to half the times recommended above. If a sensor
is damaged in any way, discontinue use immediately.
WARNING
APNEA —A pulse oximeter should NOT be used as an apnea “no
breath” monitor.
WARNING
CABLES—Route all cables away from the patient’s throat to avoid
possible strangulation.
WARNING
EARLY WARNING—A pulse oximeter should be considered an
early warning device. As a trend toward patient deoxygenation is
indicated, blood samples should be analyzed by a laboratory co-
oximeter to completely understand the patient’s condition.
WARNING
EXPLOSION HAZARD—Do not monitor SpO
2
in the presence of
flammable anesthetics or other flammable substance in combination
with air, oxygen-enriched environments, or nitrous oxide.

Table of Contents

Other manuals for GE Dash 4000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 4000 and is the answer not in the manual?

GE Dash 4000 Specifications

General IconGeneral
Resolution640 x 480 pixels
Power Supply100-240 VAC, 50/60 Hz
ECG3/5/12-lead
IBPUp to 4 channels
Battery BackupYes
CertificationsCE, UL, CSA
Parameters MonitoredECG, SpO2, NIBP, IBP, Temperature, Respiration, Cardiac Output
ConnectivityEthernet, USB, Serial
Display TypeTFT LCD
RecorderIntegrated thermal printer
Battery TypeLithium-ion
Operating Temperature0 to 40°C
Storage Temperature-20 to 60°C
Humidity Range15% to 95% non-condensing
Display Size10.4 inches
Cardiac OutputAvailable