EasyManuals Logo

GE Dash 4000 Operator's Manual

GE Dash 4000
270 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #64 background imageLoading...
Page #64 background image
4-4 Dash™ 3000/4000/5000 2000966-386D
Alarms
You can change patient status alarm levels and limits. For more information, refer to
Control Settings on page 4-5.
System Status Alarms
System status alarms are triggered by mechanical or electrical problems. They are of
lesser priority than patient status alarms.
There are three levels of system status alarms:
WARNING — Serious mechanical or electrical problems.
ADVISORY — Mechanical or electrical problems.
MESSAGE — Additional information only.
The monitors response to system status alarms is as follows.
NOTE
System status alarms cannot, in most cases, be moved from one level to another.
However, you can set default alarm levels for the following system status alarms:
ECG LEADS FAIL, SPO2 PROBE OFF, SPO2 PULSE SEARCH, NBP NO
DETERMINATION.
The System status alarm levels are configurable in MONITOR DEFAULTS >
SETUP DEFAULT DISPLAY.
Indicator CRISIS WARNING ADVISORY MESSAGE
Alarm tone Three beeps Two beeps One beep No
Alarm light Red Yellow No No
On-screen message Yes Yes Yes Yes
Automatic print
1
1
The factory default for ADULT-ICU is ON; NEONATAL-ICU and OPERATING ROOM is OFF.
Yes Yes No No
ALARM HISTORY Yes Yes Yes No
Alarm broadcast Yes Yes Yes No
Remote alarm terminal Yes Yes Yes No
Indicator WARNING ADVISORY MESSAGE
Alarm tone Repeating foghorn Single foghorn No
Alarm light Yellow No No
On-screen message Yes Yes Yes
Alarm broadcast Yes No No
Remote alarm
terminal
Yes No No

Table of Contents

Other manuals for GE Dash 4000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 4000 and is the answer not in the manual?

GE Dash 4000 Specifications

General IconGeneral
Resolution640 x 480 pixels
Power Supply100-240 VAC, 50/60 Hz
ECG3/5/12-lead
IBPUp to 4 channels
Battery BackupYes
CertificationsCE, UL, CSA
Parameters MonitoredECG, SpO2, NIBP, IBP, Temperature, Respiration, Cardiac Output
ConnectivityEthernet, USB, Serial
Display TypeTFT LCD
RecorderIntegrated thermal printer
Battery TypeLithium-ion
Operating Temperature0 to 40°C
Storage Temperature-20 to 60°C
Humidity Range15% to 95% non-condensing
Display Size10.4 inches
Cardiac OutputAvailable