Programming Site Data 1-59
March 2008
MODBUS Register Map
(cont.)
Notes:
1. Clear Totalizers: flag from the 8051 to clear totalizers in the
68332 memory.
2. Error Value: see Chapter 2, Error Codes and Screen Messages, in
the Service Manual for error codes.
3. Average:
average of channel 1 and channel 2 if both channels out of error,
channel 1 value if channel 2 is in error,
channel 2 value if channel 1 is in error,
zero
if both channels in error
4. Average Error Status:
0 = both in error
1 = chan 2 in error
2 = chan 1 in error
3 = both OK
5. Registers for Analog Inputs:
The values for the inputs apply only to special inputs (supply and
return temperature inputs are in other defined registers). The
DF868 can hold only up to 4 special inputs. Input registers for
slots and inputs that are not resident or not programmed as special
will have an output value of 0.
6. MODBUS baud rate:
5 = 2400, 6 = 4800, 7 = 9600, 8=19,200.
7. MODBUS parity:
0 = none, 1 = odd, 2 = even
8. MODBUS stop bits:
1 = 1 stop bit, 2 = 2 stop bits
9. General:
Registers are written if corresponding functions are actuated by
the user. Registers for unactuated functions are initialized to zero
at startup.

Table of Contents

Other manuals for GE DigitalFlow DF868

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE DigitalFlow DF868 and is the answer not in the manual?

GE DigitalFlow DF868 Specifications

General IconGeneral
BrandGE
ModelDigitalFlow DF868
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals